Procedure:

Screens shown in this manual represent the displays shown with a
500 psi Gauge. The BetaGauge PIR will prompt the technician for
the appropriate pressure at each calibration point.

Use the Pressure Standard to output 500.00 psi (100%). After the
output has stabilized, press the ENTER key to continue. As the
unit takes readings, the screen will show ———-. When the
readings are complete the screen should look as shown in the
illustration that follows.

Use the Pressure Standard to output 437.50 psi (87.5%). After the
output has stabilized, press the ENTER key to continue. As the
unit takes readings, the screen will show ———-. When the
readings are complete the screen should look as shown below.

Use the Pressure Standard to output 375.00 psi (75%). After the
output has stabilized, press the ENTER key to continue. As the
unit takes readings, the screen will show ———-. When the
readings are complete the screen should look as shown below.

Use the Pressure Standard to output 312.50 psi (62.5%). After the
output has stabilized, press the ENTER key to continue. As the

11

PSI

437.50

CAL MODE

PSI

375.00

CAL MODE

PSI

312.50

CAL MODE

PSI

500.00

CAL MODE