Betagauge pir serial interface instructions, List of commands – Martel Electronics BetaGauge PIR PRO User Manual

Page 17

Advertising
background image

9. BetaGauge PIR Serial Interface

Instructions

Initiating Communication

The terminal communications can be setup using terminal
communication software on a PC. The terminal settings need to
be set as follows:

• Bits per second: 9600
• Data bits: 8
• Parity: None
• Stop bits: 1
• Flow control: None
• Local echo on

List of Commands

Command

Description

CAL_START

Puts the calibrator in calibration mode

*CLS

Clears the error queue.

FAULT?

Returns an error code from the error queue

*IDN?

Identification query. Returns the manufacturer, model
number, and firmware revision level of the Calibrator.

TARE

Tares the offset pressure of the reading on the calibra-
tor

TARE?

Returns the current tare value

PRES_UNIT?

Returns the pressure unit for the upper display.

PRES_UNIT

Sets the pressure unit for the display

ZERO_MEAS

Zeros pressure of the calibrator

ZERO_MEAS?

Returns the current zero offset value

MINMAX_RST

Resets the minimum and maximum recorded values.

MIN?

Returns the minimum recorded value

MAX?

Returns the maximum recorded value

TEMP?

Returns temperature in the chosen units

CAL_STORE

Stores calibration data.

CUST_MULT?

Sets the multiplier for the custom unit type

STREAM_OFF

Turns streaming data off

STREAM_ON

Turns streaming data on

VAL?

Returns the measured pressure value in selected units

TEMP_UNIT

Used to set temperature unit

TEMP_UNIT?

Returns temperature unit

15

Advertising