unit takes readings, the screen will show ———-. When the
readings are complete the screen should look as shown below.

Use the Pressure Standard to output 250.00 psi (50%). After the
output has stabilized, press the ENTER key to continue. As the
unit takes readings, the screen will show ———-. When the
readings are complete the screen should look as shown below.

Use the Pressure Standard to output 187.50 psi (37.5%). After the
output has stabilized, press the ENTER key to continue. As the
unit takes readings, the screen will show ———-. When the
readings are complete the screen should look as shown below.

Use the Pressure Standard to output 125.00 psi (25%). After the
output has stabilized, press the ENTER key to continue. As the
unit takes readings, the screen will show ———-. When the
readings are complete the screen should look as shown below.

Use the Pressure Standard to output 62.50 psi (12.5%). After the
output has stabilized, press the ENTER key to continue. As the
unit takes readings, the screen will show ———-. When the
readings are complete the screen should look as shown below.