Martel Electronics M2000 User Manual

Page 12

Advertising
background image

ed control lines. The rest of the interface messages are sent over the data lines with the ATN
signal true. (All device-dependent and common commands are sent over the data lines with
the ATN signal false.)

An important thing to note about interface messages is that unlike device-dependent and
common commands, interface messages are not sent literally (in a direct way).

IEEE-488 standards define interface messages. Interface messages are handled automatically
in most cases.

Compound Commands

A compound command is two or more commands in a single command line. For example,
the following two commands could be entered individually,

OUT 1 V

OPER

where the M2000 sources 1 V dc, and then goes into operate, or they could be combined
into a compound command,

OUT 1 V ; OPER

using a semi-colon as a separator.

Overlapped Commands

Commands that begin execution but require slightly more time to complete are called over-
lapped commands, because they can be overlapped by the next command before they have
completed execution.

The command graphic

Overlapped shows a check for overlapped commands.

You can use the command *WAI to wait until the overlapped command has completed execu-
tion before executing the next command. For example,

OUT 1 V ; *WAI

You can also use the status commands *OPC and *OPC? to detect completion of overlapped
commands. (See "Checking M2000 Status.")

Sequential Commands

Commands that execute immediately are called sequential commands.

The command graphic

Sequential shows a check for sequential commands.

The majority of the commands are sequential.

Commands for RS-232 Only

The command graphic

IEEE-488

RS-232 indicates RS-232 interface commands.

The IEEE-488 and RS-232 interfaces both send commands to the M2000 as data, except for
those IEEE-488 functions that must be implemented as a message as specified in the IEEE-
488 standards. For example, the RS-232 interface uses the command REMOTE to place the
M2000 in the Remote mode. Although the IEEE-488 interface could also send a command
REMOTE as data, it does not because this is one of the functions that must be implemented
per IEEE-488 Standards. The relationship between these IEEE-488 messages and the equiva-
lent RS-232 emulation is shown in Table 5.

Table 5. Commands for RS-232 Only

IEEE-488 Message [1]

RS-232 Equivalent

GTL LOCAL

command

GTR REMOTE

command

LLO LOCKOUT

command

Commands for IEEE-488 Only

The command graphic

IEEE-488 indicates commands that are used for the IEEE-488

interface. This is all the commands, except for those used for RS-232 operations. (See
"Commands for RS-232 Only.") All commands are transferred over the IEEE-488 as data,

X

X

X

X

12