XV. REMOTE COMMANDS

Introduction

Remote commands duplicate activities that can be initiated from the front panel in local
operation. Following the summary table is a complete alphabetical listing of all commands
complete with protocol details. Separate headings in the alphabetical listing provide the
parameters and responses, plus an example for each command. For information on using
commands, see "Remote Operation."

Command Summary by Function

Common Commands

Command Description

*CLS

(Clear status.) Clears the ESR, the error queue, and the RQS bit in the status byte.
This command terminates pending operation complete commands (*OPC or *OPC?).

*ESE

Loads a byte into the Event Status Enable register.

*ESE?

Returns the contents of the Event Status Enable register.

*ESR?

Returns the contents of the Event Status register and clears the register.

*IDN?

Identification query. Returns the manufacturer, model number, and firmware revision
level of the Calibrator.

*OPC

Enables setting of bit 0 (OPC for "Operation Complete") in the Event Status Register to
1 when all pending device operations are complete.

*OPC?

Returns a 1 after all pending operations are complete. This command causes program
execution to pause until all operations are complete. (See also *WAI.)

*OPT?

Returns a list of the installed hardware and software options.

*RST

Resets the state of the instrument to the power-up state. This command holds off
execution of subsequent commands until it is complete. (Overlapped command.)

*SRE

Loads a byte into the Service Request Enable register (SRE).

*SRE?

Returns the byte from the Service Request Enable register.

*STB?

Returns the status byte.

*TST?

Initiates a series of self-tests, then returns a "0" for pass or a "1" for fail. If any faults are
detected, they are logged into the fault queue where they can be read by the FAULT?
query.

*WAI

Prevents further remote commands from being executed until all previous remote
commands have been executed.

Output Commands

Command Description

FUNC?

Returns the present output, measurement, or calibration function.

OPER

Activates the M2000 output if it is in standby.

OPER?

Returns the operate/standby setting.

OUT

Sets the output of the M2000.

OUT?

Returns the output amplitude of the M2000.

RANGE?

Returns the present output range. (Voltage only).

RANGELCK

Locks the present output range (Voltage only).

RANGELCK?

Returns the RANGELOCK state (Voltage only).

STBY

Deactivates the M2000 output if it is in Operate mode.