Error code listing, Rs-232 port commands, Status commands – Martel Electronics M2000 User Manual

Page 20: Remote command listing

Advertising
background image

RS-232 Port Commands

Command Description

LOCAL

Puts the M2000 into the local state and disables lockout.

LOCKOUT

Puts the M2000 into the lockout state. This command duplicates the

IEEE-488 LLO

(Local Lockout) message.

REMOTE

Puts the M2000 into the remote state. This command duplicates the

IEEE-488 REN

(Remote Enable) message.

Status Commands

Command Description

FAULT?

Returns the most recent error code in the M2000 error queue, then removes that error
code from the queue.

Error Code Listing

Error

Message

Number

Class Description

1

DDE

Error queue overflow.

101

CME

A non-numeric entry was received in a field that should contain a numeric
entry.

102

EXE

The numeric field exceeds 10 characters.

105

EXE

Entry is above upper limit for the selected output range.

106

EXE

Entry is below lower limit for the selected output range.

107

EXE

OUT? was received when the M2000 was not in a Source mode.

108

CME

A required command parameter was missing.

110

CME

An invalid RANGELCK parameter (not ON or OFF) was received.

111

EXE

RANGELCK ON was received when the M2000 is not in Volts mode.

117

CME

An unrecognizable command was received.

118

CME

An invalid parameter was received.

120

EXE

The serial input buffer overflowed

121

EXE

The command string buffer overflowed.

122

QYE

The serial output buffer overflowed.

123

DDE

The output overloaded.

124

DDE

M2000 is out of tolerance. This error is set after a failed initialization or a
failed *TST?.

125

DDE

M2000 has a ADC failure. This error is set after a failed initialization or a
failed *TST?.

Remote Command Listing

The following is an alphabetical list of all M2000 commands and queries, including common
commands and device-dependent commands. Each command title includes a graphic that
indicates remote interface applicability, IEEE-488 and RS-232, and command group:
Sequential and Overlapped.

*CLS

IEEE-488

RS-232

Sequential

Overlapped

(Clear Status command) Clears the ESR, the error queue, and the RQS bit in the status byte.
This command terminates pending operation complete commands (*OPC or *OPC?).

Parameter: (None)

Example: *CLS

Clear the ESR, the error queue, and the RQS bit in the status byte.

X

X

X

20