6.4.3 1 Cycle Mode (SIg)

Contents

1 Cycle (Compression. Expansion) Movement is performed.

Between the Limits, the stand repeats movement for 1 time cycle only. The Stand will proceed towards the Limit
Knob in the PUSH/PULL direction. It will hit that Limit, then move in the opposite direction all the way to the
opposing Limit Knob.

If either PUSH Key or PULL Key is pressed, the stand starts moving toward that respective direction. STOP
key can instantly end the test.

SIG

SIG

Upper Limit Knob.

Lower Limit Knob.

Stops by position of Adjustable Limit Knob.
After stopping for 1 second, it turns back and moves
towards PUSH Direction.

Stops by Limit Switch in ON position.

PULL

LMT

LMT

PULL Speed

PUSH Speed

SIG

SIG

Upper Limit Knob.

Lower Limit Knob.

Stops by Limit Knob position.
After stopping for 1 second, it turns back and moves
towards PULL Direction.

Stops by Limit Knob position.

PUSH

LMT

LMT

PULL Speed

PUSH Speed

Change speed by UP, DOWN Key. SIG Mode and CON Mode Speed Setting are commonly used. During
movement, Speed variation is not possible. When no movement is occurring (at idle), the display shows
movement Mode and the latest operated speed. When the movement starts, the respective memorized
movement speed is displayed.