9. Trouble Shooting

Following is the explanation of the examples of general troubles and respective solutions. If you still have issues,
then please enquire with your local agent or contact our customer service.

Confirm whether AC100~240V Power Supply is provided to the Power Supply Cord.

Phenomenon.

Trouble Solution Method.

Even though Power Supply is ON, the LCD
does not display.

Supply AC100 ~ 240 V

In case it is interrupted, confirm whether any abnormalities exist in the
Power Supply, replace the Fuse.
For Fuse Replacement, use FGMB 250V 5A PBF Fuji Terminal
(Appropriate goods)

Confirm whether the Stand Back Side Fuse is interrupted.

Arm / Alarm do not work.

"ALM" lights up at LCD Display.
Motor Alarm

"EMg" lights up at LCD Display.
(Emergency Stop Switch ON)

"OVR" lights up at LCD Display.
(Overload detection)

"LMT" lights up at LCD Display.
(Upper or Lower Limit knob
detection).

Confirm whether excessive load exists.

Confirm whether Emergency Stop Switch is ON.

Confirm the Over Load of Force Gauge.

Reaches to Limit Switch.

Remove Stand Load

After confirming that there is no abnormal condition and confirming that
there is no problem, then put the Emergency Stop Switch to OFF.

Remove the excessive load of Force Gauge.

Change the position of Limit Switch.