LMT

EMg

7. Display Error

OVR

OVR

ALM

Lower Limit Knob is reached. Upper Display, LMT
and MODE flash alternately. Main Display becomes
PUSH.

Upper Limit Knob is reached. Upper Display, LMT
and MODE flash alternately. Main Display becomes
PULL.

The Force Gauge's sensor PUSH over load was
detected. Upper Display alternately lights-up with
OVR and MODE.

The Force Gauge's sensor PULL over load was
detected. Upper Display alternately lights-up with
OVR and MODE.

Emergency Stop Switch was pressed.
Upper display alternately lights-up with EMG and
MODE.

A motor abnormality was generated.
Upper display flashes ALM.

In normal Limit movement, this Display is generated.

In normal Limit movement, this Display is generated.

Release opens the movement and load. Confirm the
load given to the Force Gauge and confirm whether
the rated capacity of the installed Force Gauge is
appropriate.

After avoiding the abnormal condition, put the
Emergency Stop Switch to OFF.

Switch the Power Supply OFF for 1 minute, then
turn ON. If again the same Error is generated,
then enquire with your local agent or contact our
customer service.

Release opens the movement and load. Confirm the
load given to the Force Gauge and confirm whether
the rated capacity of the installed Force Gauge is
appropriate.