Techne TC-3000G User Manual

Page 23

Advertising
background image

0.5ML TUBES

In order to fit the maximum number of tubes in the block, they need to be placed so that all the hinges
are in the same relative orientation. An example is illustrated below.

PLATES AND SEALS

The 48 x 0.2ml block of the TC-3000G will accept 48-well plates as well as individual and strip 0.2ml
tubes. To avoid the possibility of sample evaporation Bibby Scientific Ltd recommends that plates are
sealed with appropriate caps or heat seals. We do not recommend the use of adhesive seals, but if their
use is unavoidable then the supplied silicone rubber mat should be placed on top of the sealing film to
ensure good thermal contact with the heated lid.

Place the tubes in the block in the same
orientation in order to achieve the maximum
capacity.

Always make sure the pressure on the heated
lid is balanced by spreading the tubes across
the block or inserting some dummy tubes if
necessary.

When you switch on the unit a screen similar
to the one shown here will appear briefly. This
indicates the software version and block type
installed in the unit.

When the initial screen has disappeared the
screen shown here will automatically appear.
There are four options:

1. 'Programs' for creating and running

programs;

2. 'Information' for obtaining information

about the unit e.g. serial number;

3. 'Configuration' for setting up program

defaults and settings.

4. Gradient calculator for displaying column

temperatures of a set gradient.

Each of these is explained in detail in the
following sections.

TC-3000G

Version

x.xx

Block

48x0.2ml GRAD

Programs

[0]

Information

Configuration

Gradient calculator

[ , to select

]

SWITCHING ON

21

Advertising