2. Before connecting the instrument to the mains electricity supply, check the voltage against the rating

plate (located on the back of the unit). Please note that the unit must be earthed to ensure proper
electrical safety. The TC-3000G has a switch-mode power supply which is suitable for use with
universal AC input voltages between 100 and 230V. Connect the mains cable to a suitable plug
according to the table below.

Connections

220V-240V, 50/60Hz Supply

110V-120V, 50/60Hz Supply

Live

brown

black

Neutral

blue

white

Earth

green/yellow

green

3. UK ONLY: The fused plug supplied with the mains cable is fitted with a 10 Amp fuse to protect the

instrument and the user.

4. Plug the mains cable into the socket on the back of the instrument.

5. Place the unit on a suitable flat bench or in a fume cupboard if required, ensuring that the air inlet

vents on the underside are free from obstruction. Position the instrument with a minimum distance all
round of 200mm from walls or other items and between each unit if multiple units are being used to
permit efficient air flow for each instrument.

6. Switch on the instrument:

I

mains switch On

O mains switch Off

Replacement cable

Should the mains lead need replacement, a cable of 1mm

2

of harmonized code H05VV-F connected to

an IEC320 plug should be used. IF IN DOUBT CONSULT A QUALIFIED ELECTRICIAN.

WORKING CONDITIONS

The thermal cycler is designed operate under the following conditions:

• Indoor use
• Ambient temperature range +5°C to +40°C
• Altitude to 2000m
• Up to 95% relative humidity, non-condensing
• Mains supply fluctuations not exceeding 10%
• Over voltage category II IEC 60364-4-443
• Pollution degree 2

Note: The control specifications are quoted at an ambient temperature of 20°C. The specification may
deteriorate outside an ambient temperature of 10°C to 30°C.

The instrument has been tested for radio frequency interference and is certified under EN61326.