Thermo Fisher Scientific CyberScan pH 300/310 User Manual

Page 33

Advertising
background image

30

P4.1: READY Indicator and auto endpoint function

Program P4.1 lets you select “READY indicator on” to
indicate when the reading is stable*, or select “READY
indicator off” for faster meter response.

pH 310 meter only

: Program P4.1 also lets you switch the

Auto endpoint function on or off. Select auto endpoint on to
“hold” the reading when it is stable for more than 5 seconds.
The display automatically freezes, and the HOLD indicator
appears on the left side of the display. Press the HOLD key
to release the display and access other functions. Select
auto endpoint off to deactivate this feature.

From measurement mode

1.

Press SETUP key to enter Set Up mode.

2. Press the MI/▲ or MR/▼ keys to scroll through

subgroups until you view parameter P4.0.

3.

Press the ENTER key to select parameter 4.1.

4. Press the MI/▲ or MR/▼ keys to select the

configuration you require.

OFF switches the READY indicatory off.

ON switches the READY indicator on.

pH 310 meter only

: ON and HOLD together switches

the auto endpoint feature on.

5. Press the ENTER key to confirm selection and to

proceed to step 4 of P4.2. Press the CAL/MEAS key to
return to measurement mode.

NOTES:

Meter default is set for READY indicator on and auto endpoint function off.

*The READY indicator appears and the reading holds until the measured value exceeds the tolerance (

± 0.02 pH; ±

0.8 mV <400 mV;

± 1.2 mV > 400). Then the READY annunciator turns off.

Figure 29 - P4.1: Selecting READY

function)