7.6

P5.0: Resetting to factory default settings

Available in pH 300 meter only

In the pH 300 meter, Program 5 lets you reset all
parameters to factory default settings. This clears all
calibration data, memory, and any other setup functions you
might have changed.

From measurement mode

1.

Press SETUP key to enter Set Up mode.

2. Press the MI/▲ or MR/▼ keys to scroll through

subgroups until you view parameter P5.0.

3. Press the ENTER key to enter parameter P5.0. See

Figure 33.

4. Press

the

MI/▲ or MR/▼ keys to toggle between NO

and YES.

NO retains current settings

YES resets to factory default settings.

5. Press the ENTER key to confirm selection and to

return to the measurement mode. Otherwise press
CAL/MEAS key to return to measurement mode
without resetting to factory default.

Figure 33 - P5.0: Reset to factory

default values (pH 300 meter only)