Thermo Fisher Scientific Eutech CON 700 User Manual
Page 14
10
Press
or
to choose READY “On”, READY “OFF”, or
Auto HOLd.
Press
to confirm.
6.5 P3.2 ºC ºF (Celcius Or Fahrenheit)
Press
.
Press
or
to select ºC or ºF.
Press
to confirm.
6.6 P3.3 AtC (Automatic Temperature Compensation)
Press
.
Press
or
to select “Yes” or “No”.
Press
to confirm.
6.7 P3.4 tdS (Total Dissolved Solids factor)
Press
.
Press
or
to select the desired TDS factor (.40 to
1.00).
Press
to confirm.
6.8 P3.5 t.CO (Temperature Coefficient)
The temperature coefficient is the amount of change in conductivity per
degree temperature (% per ºC). The CON 700 is factory set to a
temperature coefficient of 2.1 % per ºC. For most applications this will
provide good results. The meter allows adjustment from 0.0 to 10.0.
TIP: Select 0.0% for uncompensated measurements. The temperature
will be measured by the electrode and displayed in measurement
mode—without compensation.