11

See Section 8

– Calculating Temperature Coefficients.

Press

.

Press

or

to select the desired value (0.00 to 10.0).

Press

to confirm.

6.9 P3.6 t.nr (Normalization Temperature in ºC)

When Automatic Temperature Compensation is used, measurements
are adjusted by the temperature coefficient to the normalization
temperature. The default value is 25 ºC.

Press

.

Press

or

to select the desired value (15.0 to 30.0).

Press

to confirm.

6.10 P3.7 ACAL (Automatic Conductivity Calibration)

See Section 4.1 for more details on Automatic Conductivity Calibration.

Press

.

Press

or

to select “Yes” (Automatic) or “No” (Manual).

Press

to confirm.

6.11 P3.8 SPC (Single Point Calibration)

See Section 4.2 for details on Single Point & Multi Point Calibration.

Press

.

Press

or

to select “Yes” (Single) or “No” (Multi).

Press

to confirm.