Electrode information, Calibration, 1 automatic or manual calibration – Thermo Fisher Scientific Eutech CON 700 User Manual
Page 8
 
4
3. ELECTRODE
INFORMATION
The CON700 includes an electrode with a nominal cell constant of 
k = 1.0, built-in temperature sensor, and 1 meter cable. The Ultem 
body housing has good chemical resistant properties. The 
electrode design offers fast temperature response and reduces air 
entrapment, ensuring accurate, repeatable, and stable readings. 
The wetted materials of the probe include:
•
Polyetherimide (Ultem) – protective probe guard
•
Polybutylterphalate (Valox) – sensor housing
•
Stainless Steel (SS 304) – 2 bands
The protective probe guard can be removed temporarily for 
cleaning however it must be re-attached during measurement and 
calibration. Erroneous results will occur while the probe guard 
is removed. 
Always immerse the probe beyond upper steel band for best 
results. Use the fill line on the outside of the probe guard for 
reference. 
4.
CALIBRATION
4.1 Automatic or Manual Calibration
The CON 700 is capable of automatic or manual calibration for 
conductivity, and manual calibration for TDS. 
In the automatic calibration mode, the meter will automatically select 
one of (4) conductivity calibration standard values depending on the 
range and normalization temperature being used (TABLE 1). 
If you only use calibration standards that are listed in TABLE 1, 
automatic calibration is recommended. If you intend to calibrate with 
one or more standards that are not listed in TABLE 1, the CON 700 
must be set for manual calibration instead.