Thermo Fisher Scientific Eutech CON 700 User Manual

Page 16

Advertising
background image

12

6.12 P3.9 CELL (Nominal Cell Constant)

The CON 700 includes a probe with a nominal cell constant (k) of 1.0.
Use probes with k = 0.1 and 10 (sold separately) for improved
performance in extreme sample ranges. Use this setup function to
change the cell constant if necessary. Meter default is 1.0 to match the
included probe.
k = 0.1 ideal for low measurements <20 µS (<10 ppm).
k = 1.0 ideal for mid-range measurements
k = 10 ideal for high measurements >20 mS (>10 ppt).

Press

.

Press

or

to select 0.1, 1.0, or 10.0

Press

to confirm.

6.13 P4.0 rSt (Reset)

Press

.

Press

or

to select “Yes” (Reset) or “No” (Cancel).

If “Yes”, press

or

to select “Cal” (calibration reset

only) or “FCt” (complete reset to factory default settings).

Press

to confirm.

6.14 P5.0 CLr (Clear Memory)

Press

.

Press

or

to select “Yes” (Erase memory) or “No”.

Press

to confirm.