Thermo Fisher Scientific Eutech CON 700 User Manual
Page 16
12
6.12 P3.9 CELL (Nominal Cell Constant)
The CON 700 includes a probe with a nominal cell constant (k) of 1.0.
Use probes with k = 0.1 and 10 (sold separately) for improved
performance in extreme sample ranges. Use this setup function to
change the cell constant if necessary. Meter default is 1.0 to match the
included probe.
k = 0.1 ideal for low measurements <20 µS (<10 ppm).
k = 1.0 ideal for mid-range measurements
k = 10 ideal for high measurements >20 mS (>10 ppt).
Press
.
Press
or
to select 0.1, 1.0, or 10.0
Press
to confirm.
6.13 P4.0 rSt (Reset)
Press
.
Press
or
to select “Yes” (Reset) or “No” (Cancel).
If “Yes”, press
or
to select “Cal” (calibration reset
only) or “FCt” (complete reset to factory default settings).
Press
to confirm.
6.14 P5.0 CLr (Clear Memory)
Press
.
Press
or
to select “Yes” (Erase memory) or “No”.
Press
to confirm.