Relative Humidity:

20% - 95% non-condensing

Source Resistance:

1000

maximum (thermocouple)

Lead Resistance:

50

per lead maximum balanced (Pt100)

Performance Under Operating Conditions

Temperature Stability:

0.005% of span/

°

C change in ambient

temperature.

Cold Junction

Compensation

(thermocouple only):

Better than

±

1

°

C.

A.10

ENVIRONMENTAL

EMI Susceptibility:

Complies with EN50082-1:1992,

EN50082-1:1995. NOTE: For RF

electromagnetic fields (10V/m 80% AM

1kHz), the reading accuracy may be

impaired by up to 3oC in the frequency

band 273 to 470MHz and by up to 1.4oC

in the frequency band 785 to 1000MHz.

EMI Emmissions:

Complies with EN50081-1:1992,

EN50081-2:1993.

Safety:

Complies with EN61010-1.1993.

Supply Voltage:

90 - 264V, 50/60Hz (standard)

20 - 50V AC/DC (option)

Power Consumption:

4watts approximately

Front Panel Sealing:

To IP66 (Nema 4)

A.11

PHYSICAL

Dimensions:

Height - 48mm

Width - 96mm

Depth - 100mm

Mounting:

Panel-mount; press-fit fixing strap supplied.

Panel cut-out - 45mm x 92mm

Terminals:

Screw type; combination head.

Weight:

0.21kg maximum.