Installation & Operation Manual

INTRODUCTION

Used in conjunction with any Badger Meter Impeller flow monitor or endpoint, Badger Meter non-magnetic Impeller flow

sensors provide an accurate reading of the rate of liquid flow as well as total accumulated flow. A number of sensor models

are offered, which cover applications for a wide range of pipe sizes and pressure/temperature specifications.
The flow sensors generate a frequency which is proportional to flow rate. An internal preamplifier allows the pulse signal to

travel up to 2000 feet without further amplification. Power to operate the sensor is provided by the flow monitor. The impeller

bearing assembly, shaft and o-rings are replaceable in the field.
Badger Meter Impeller flow sensors feature a closed, six-bladed impeller design, using a proprietary, non-magnetic sensing

technology. The forward-swept impeller shape provides higher, more constant torque than four-bladed impeller designs,

and is less prone to fouling by water-borne debris. The forward-curved shape, coupled with the absence of magnetic

drag, provides improved operation and repeatability, even at lower flow rates. As the liquid flow turns the impeller, a low

impedance signal is transmitted with a frequency proportional to the flow rate.
Sensors of similar type are interchangeable, so there is no need for recalibration after servicing or replacement.

ELECTRONIC TYPES

Badger Meter provides several basic sensor configurations using the same impeller element. This allows for a wide range of

applications and pipe sizes. Sensors are normally supplied with 20 feet of 2-conductor 20 AWG shielded U.L. type PTLC

105° C cable. Optional sensors designated with the prefix "IR" feature two single conductor 18 AWG solid copper wire leads

48 inches in length with U.L. Style 116666 direct burial insulation. These IR models are used in below grade applications

such as irrigation, municipal, and groundwater monitoring. All Badger Meter Series 200 sensor electrical components are

self-contained. Pressure/temperature ratings for the various models are contained in "SPECIFICATIONS" on page 10. These

models can be further described as follows.

Standard Sensor

Designed for indoor or protected area applications such as HVAC, pump control, and industrial process monitoring where the

flow rates are between 0.5…30 feet/second and temperatures are below 221° F. Standard sensors are supplied with 20 feet of

2-conductor 20 AWG shielded U.L. type PTLC 105° C cable.

IR Sensor

Designed for below grade applications such as irrigation, municipal, and groundwater monitoring where the flow rates are

between 0.5…30 feet/second and temperatures are below 180° F. IR sensors are supplied with two single-conductor, 18 AWG

solid copper wire leads 48 inches in length with U.L. Style 116666 direct burial insulation.

CSA Sensor

Designed for indoor or protected area applications where intrinsic safety is required and the flow rates are between

0.5…30 feet/second and temperatures are below 140° F. CSA sensors are supplied with 20 feet of 2-conductor 20 AWG

shielded U.L. type PTLC 105° C cable. These sensors must be used with an approved safety barrier.

Model 228PV (Formerly 220P)

These models feature a modified PVC tee with solvent weld socket end connections, and a removable, PPS or PVDF sensor

insert. Sizes of 1-1/2", 2", 3" and 4" are available.

The impeller style flow sensor described in this manual is not intended for use in safety critical applications.

Use of the device in this manner is done at the sole discretion of the customer and/or end user of the device
The impeller style flow sensor described in this manual is not intended for use in systems with flammable liquids or

gases. Additionally, the device is not intended for systems containing hazardous fluids, or fluids other than water.
The impeller style flow sensor described in this manual must be installed in accordance with all local and federal codes

or end-use standards, as applicable.
If the devices described in this manual are used in a manner not specified by the manufacturer, the protection provided

by the equipment may be impaired.