Badger Meter Water Conditioning User Manual

Page 11

Advertising
background image

Figure 4: Turbo Meters Serviced Without Removal from the Line

EXERCISE PRECAUTION TO MAKE CERTAIN THAT FLUID DOES NOT SPRAY FROM THE METER. THE LIQUID SPRAY COULD

CONTACT ELECTRICAL EQUIPMENT AND CREATE AN ELECTRICAL HAZARD.
4 Allow the meter to drain and relieve internal pressure
5 When pressure is relieved, remove the head bolts and shift the head assembly toward the outlet-end of the meter to

release the head-to-housing seal Then lift the entire head assembly from the housing and, at the same time, tilt the

operating head up to prevent the rotor from falling off the rotor spindle

NOTE:

N

Rotors are furnished with two-pole or four-pole magnets depending on the type of accessory

6 When the operating head has been removed, lift the rotor out of the rotor cage and set it aside Remove the head gasket

NOTE:

N

A dummy cover plate is available to close the meter in the line and continue service while the operating head

assembly is being inspected or parts are being replaced

7 To remove the register base from the head assembly, loosen the seal screw on the base with a screwdriver Rotate the base

45 degrees in either direction to release it from the bayonet lock detents on the head assembly

8 The register base retaining ring is accessible through the rotor cage in the head assembly With a Truarc right-angle pliers

#1549, expand and remove the register retaining ring on the extension tube When this is accomplished, the register base

can be removed from the head assembly

9 To reinstall or replace the register base, repeat the procedure above in reverse making sure that the retaining ring has

been reinstalled on the extension tube

NOTE:

N

The head unit has a gasket bonded to the periphery of the rotor cage Depending on the chemical solution to be

metered, the gasket will be made of EPR, Buna N or Viton A material Re-inserting the head assembly into the meter

housing correctly will require compressing the gasket slightly This is done by tilting the top of the register base

on the head toward the inlet side of the meter The meter head has a pilot diameter machined into the cage at the

gasket face This pilot diameter must extend through the housing gasket inner diameter and into the housing bore

After lowering the head assembly into the housing, move the entire head assembly straight toward the inlet side of

the meter The pilot diameter must snap into the bore of the housing to provide a tight seal

10 With the head gasket aligned and the head assembly properly positioned in the housing, reinstall and tighten the head

bolts (between 90…180 in-lb of torque)

11 Close the flushing system drain valve Open the upstream valve partially, then open the downstream valve slightly which

will purge any air from the service line Then open both valves completely

User Manual

Page 11