Series BPI/RTD Battery Powered Temperature Indicator

Specifications - Installation and Operating Instructions

Bulletin E-90-BPI/RTD

DWYER INSTRUMENTS, INC.

Phone: 219/879-8000

www.dwyer-inst.com

P.O. BOX 373 • MICHIGAN CITY, INDIANA 46361, U.S.A. Fax: 219/872-9057

e-mail: [email protected]

The Series BPI is a range of battery powered temperature

indicators. The Series BPI-XX1 accepts an RTD sensor; the

Series BPI-XX2 accepts a thermocouple sensor. The sensor

type, temperature units and other flexible configuration settings

are selectable via a simple to use menu system, which is

navigated by the use of three push-button keys located on the

rear of the display. The entire assembly is sealed into a cap that

fits directly onto the connecting heads.

Please inspect the packaging and instrument thoroughly for any

signs of transit damage. If the instrument has been damaged,

please notify your shipper immediately.

INSTALLATION

Configuration - Fit the battery as described in Battery section,

then follow the simple configuration menus printed on the

reverse side of this sheet. This procedure can be performed with

or without the sensor connected. Fit the correct units legend to

the front panel face.

Enclosure - The case offers various options for sensor entry.

Please refer to the data sheet for further details. The sensor

may be fixed directly to the housing, or remotely connected

using cable and gland. Cable length must be no greater than 20

meters to comply with CE certification. All external

cabling/sensor entries must maintain IP67 rating.

Wiring

BPI/RTD Connection

SPECIFICATION @ 68°F (20°C)

Inputs: Thermocouple or RTD depending on model.

Accuracy: Thermocouple input: ±0.1% FS, ±0.5°C

(plus sensor); RTD: ±0.2°C ±0.1% of reading (plus sensor error).

Power Requirements: 3.6 V Lithium battery.

Battery Life: >2 years.

Display: 4-digit LCD.

Resolution: 0.1°C.

Ambient Operating Temperature: 14 to 158°F (-10 to 70°C).

Storage Temperature: -4 to 185°F (-20 to 85°C).

Weight: 6.0 oz (170 g).

Front Panel Protection: NEMA 4X (IP65).

Agency Approval: CE.

Battery

To install battery, first gain access to the battery by unscrewing

the cap retaining nut and removing cap from base. To remove

old battery, ease out the positive end of the battery from the

holder, using a screw driver blade. To fit battery, insert battery

negative into the spring contact end of the battery holder, then

press battery into place.

The units monitor their batteries. When the battery approaches

the end of its lifetime (over 2 years under normal conditions), the

display will toggle the message “LO BAT”.

Please observe the battery warning.

Ensure the instrument is installed with adequate protection

against the environment. IP67 rating must be maintained.

Try to avoid installing the equipment close to sources of extreme

temperature and electrical or electromagnetic interference.

Any cleaning of the instrument should be carried out using a mild

detergent and a soft cloth. No solvents or abrasive cleaner

should be used.

+

BH1

DEC

INC

CYC

RTD
Connection

13/16

[20.7]

13/64

[5]

Ш2-5/8

[Ш66.7]

Ш2-7/8

[Ш73]

Fire, Explosion and severe burn hazard, DO NOT

Recharge, Crush, Disassemble, Heat above

100°C (212°F), Incinerate or expose contents to water. NEVER

short circuit the battery. Disposal, the battery may be regulated

by national and local regulations. Please follow the instructions

of the proper regulator.