Dwyer ISDL User Manual

Dwyer instruments, inc, Series isdl temperature/humidity data logger

Advertising
background image

Bulletin DA-ISDL

DWYER INSTRUMENTS, INC.

Phone: 219/879-8000

www.dwyer-inst.com

P.O. BOX 373 • MICHIGAN CITY, INDIANA 46361, U.S.A.

Fax: 219/872-9057

e-mail: [email protected]

Specifications - Installation and Operating Instructions

Series ISDL Temperature/Humidity Data Logger

The Series ISDL Temperature/Humidity Data Logger accurately records
temperatures that range from -40 to 185°F (-40 to 80°C). The easy to use
DL700 software makes retrieving data simple and allows for user
selectable sampling rates. The ISDL can easily be started and stopped
from a PC or delayed to start up to six months in advance.

Before the logger is submerged, screw on all of the cap(s) tightly to make
sure that the logger is properly sealed.

SOFTWARE INSTALLATION
1. Insert installation CD and click Install Dwyer 2.00.
2. Follow onscreen instructions.
3. Next, click Install USB Interface Drivers on the CD Contents screen.
4. Click Install.

A message may appear stating that the software has not passed
Windows Logo Testing. This software has been tested and works
properly. Click Continue Anyway if this window appears.

SPECIFICATIONS
Range:
Temperature: -40 to 176°F (-40 to 80°C); Humidity: 0 to 100%RH.
Memory Size: ISDL-T10: 32,767 readings; ISDL-R10: 21,845 readings
per channel.
Accuracy: Temperature: ±0.9°F (±0.5°C); Humidity: ±3%RH.
Resolution: Temperature: 0.2°F (0.1°C); Humidity: 0.5%RH.
Sampling Method: Stop on memory full.
Sampling Rate: Selectable from 2 sec to 12 hrs.
Computer Requirements: Windows® 95, Windows® 98, Windows®
2000, Windows® ME, Windows NT®, and Windows® XP operating
system, one free USB port.
Power Requirements: User replaceable, 3.6V lithium battery.
Battery Life: 1 year (approx).
Interface: USB port (interface cable required).
Material: Aluminum or stainless steel (depending on model).
Weight: ISDL-T10: 4 oz (110 g); ISDL-T10-SS: 8 oz (230 g); ISDL-R10: 5
oz (145 g); ISDL-R10-SS: 10 oz (285 g).
Agency Approvals: CE.

Windows® and Windows NT® are registered trademarks of Microsoft.

NOTICE

NOTICE

3-53/64 [97.11]

ø1[25.40]

ISDL-T10

ISDL-R10

VENTS TO

HUMIDITY SENSOR

4-63/64

[126.75]

ø1 [25.40]

DA-ISDL:TEMPLATE 5/6/09 11:44 AM Page 1