FUTEK IPM52x (D520) Digital Display User Manual

Page 27

Advertising
background image

26

C.

"RESET" assigns

F key for resetting latched outputs

Pushing

F key will reset all relays in an alarm condition.

D.

"SCALE2" assigns

F key to initiate 2

nd

scaling to be applied.

Pushing

F key will switch displayed channel to 2

nd

scaling.

Front panel S2 LED will light to indicate 2

nd

scaling is

applied.

Push

F key while S2 is selected will change scaling back

to initial scaling

If different channel is selected while 2

nd

scaling is

selected, it will also use the 2

nd

scale factor applied to it.

E.

"SHUNT" assigns F key to "SHUNT" Cal function.

If assigned to display calibration, pressing

F key for 3

seconds will initiate shunt cal relay to close. This

displays calibration value.

Note: input should be unloaded.

F.

"NET" assigns F key to switch displayed channel between gross

and net

Pushing

F key will switch displayed channel to display

net.

Front panel NET LED will light to indicate NET is being

displayed.

Push

F key while net is selected will change display back

to gross.

If different channel is selected while net is selected, it

will also display net.

If assigned to do a "SHUNT" calibration, pressing

F key

triggers display of CAL with selection of "YES or NO".

"NO" aborts and returns to Run mode.

"YES" initiates the shunt calibration procedure.

G.

"PEAK 1" and "PEAK 2" assigns programmed peak function to

F key