In-e-hl-nt-v2_12 – ROTRONIC HygroLog HL-NT User Manual

Page 15

Advertising
background image

IN-E-HL-NT-V2_12

Rotronic AG
Bassersdorf, Switzerland

Document code

Unit

HygroLog HL-NT data logger:
instruction manual

Instruction Manual

Document Type

Page 15 of 48

Document title

© 2009-2012; Rotronic AG IN-E-HL-NT-V2_12

Event

Description

New memory card
inserted

The memory card was replaced while data was being logged. The data
will be split between two different memory cards. Possibly, some data
has been lost.

Humidity adjusted

Humidity adjustment of the probe connected to input #

Temperature adjusted

Temperature adjustment of the probe connected to input #

Logging started

Start recording data from input #

Logging stopped
manually

Data recording of input # ended before the programmed stop time

Logging ended
automatically

Data recording of input # ended at the programmed stop time

Out-of-limits value
detected

An out-of-limits value was newly detected on input #

Battery almost empty

Battery voltage dropped below 6.5V

Battery empty

The battery is empty and the logger has powered itself off (keeping
power up could result in erroneous data or loss of data)

Beginning accumulator
charge

Starting to charge the rechargeable battery

Accumulator charge
ended

Rechargeable battery full

MFG command

Reserved for the factory

Lost data, memory card
not ready

While recording data, the logger could not find the memory card. The
data could not be written to the EEPROM, and was lost

HygroClip probe
connected

1

A HygroClip probe was connected to input #

HygroClip probe
disconnected

1

A HygroClip probe was removed from input # or the input can no
longer communicate with the probe

External power
connected

The A/C adapter was connected and is being powered

External power
removed or faulty

The A/C adapter was disconnected, or failed, or is not being powered

Device time changed /
adjusted

Device date and time prior and after adjustment (up to firmware v1.1d
only)

Prior device time

Device date and time prior to change (firmware v1.2a and up)

New device time

Device date and time after change (firmware v1.2a and up)

RS-485 address
changed

The RS-485 address was changed

EEPROM erased

Reserved for the factory

Docking station
disconnected

2

The docking station was disconnected or there is no longer any
communication with it

Advertising