Rs-485 multi-drop network, In-e-hl-nt-v2_12, 3 rs-485 multi-drop network – ROTRONIC HygroLog HL-NT User Manual

Page 38

Advertising
background image

IN-E-HL-NT-V2_12

Rotronic AG
Bassersdorf, Switzerland

Document code

Unit

HygroLog HL-NT data logger:
instruction manual

Instruction Manual

Document Type

Page 38 of 48

Document title

© 2009-2012; Rotronic AG IN-E-HL-NT-V2_12

Docking station models with an Ethernet (TCP/IP) interface are shipped with a Device
Configuration Certificate that provides information about the factory configuration settings of both
the Ethernet module. See example below:

Setting

Value

DHCP or Static Address

Static

IP Address

192.168.1.1

Subnet Mask

255.255.255.0

Default Gateway

192.168.1.0

TCP Port

2101

MAC Address

00:40:9D:28:2D:2A

Firmware Release (device)

1.4

User Name (web interface)

rotronic

Password (web interface)

wlan

Serial Port Configuration Profile

TCP Sockets

Basic Serial Settings

Baud Rate

57600

Data Bits

8

Parity

None

Stop Bits

1

Flow Control

None

9.3 RS-485 multi-drop network

Most models of docking station feature a RS-485 connector. When the number of available PC
ports is limited, this connector can be used to connect together up to 64 devices in a multi-
dropped arrangement. In principle, an unlimited number of such networks can be monitored with
the HW4 software, but each RS-485 multi-drop network is limited to 64 devices. A docking station
with HygroLog HL-NT can be used either as a slave or a master, without special configuration.
The master is automatically the device that is directly connected to the PC or LAN by means of a
RS-232 port, USB port or TCP/IP port.

RS-485 Address: the RS-485 address is used to identify the individual devices that are
connected to a RS-485 multi-drop and must be unique to each device. This address is part of the
HygroLog HL-NT communications protocol. The factory default for the RS-485 address is 00. In
general, this address should not be changed manually (see note below). As far as HW4 is
concerned, all masters can share the same RS-

485 address. When the HW4 command “Search

for RS-

485 slaves” is being used, HW4 automatically changes from 00 to 01 the RS-485 address

of any master that has slaves attached. As a result, all masters end up with the same RS-485
network address (01). In addition, HW4 automatically changes the RS-485 address of the slaves
in each multi-drop from 00 (factory default) to a unique address ranging from 02 to 64. The same
address range (02 to 64) is used again when there is more than one multi-drop RS485 network.

Notes:

- We recommend using HW4 to give each HygroLog HL-NT a unique name for easier

identification.

- In the situation where an RS-485 network is to be monitored by software other than HW4,

you will have to use HW4 to give each device a unique RS-485 address prior to connecting to
the network.