9. Operation on a network

9.1 Baud rate and communications protocol compatibility

requirements


Depending on the model, devices from ROTRONIC use one of the following Baud rates as the
factory default:

57600 bps: HygroLog HL-NT data logger and docking stations

19200 bps: HygroPalm and HygroLab indicators, HygroFlex and M33 transmitters,
HygroClip Alarm programmable alarm card, HygroClip DI interface, HygroStat MB Thermo-
Hygrostat


IMPORTANT:

All devices connected to the same RS-485 multi-drop network (master and slaves) must
use the same Baud rate and the same communications protocol.

The communications protocol used by the HygroLog HL-NT is compatible with previous
versions of the HygroLog NT and with legacy products such as the HygroFlex 2 or 3. The
HygroLog HL-NT cannot be used in the same RS-485 multi-drop network as a HF4 or HF5
transmitter and similar products that rely on the RO-ASCII communications protocol..

In the case of a docking station with an internal Ethernet (TCP/IP) module, the Baud rate of
the Ethernet module must be the same as the Baud rate of the HygroLog HL-NT used with
the docking station.


Devices and internal Ethernet modules with mismatched Baud rate will not communicate
together.

The Baud rate configuration of a ROTRONIC device can be changed with the Device
Manager
function (Digital Interface tab) available from within the HW4 software.

The Baud rate of a Digi internal Ethernet module can be changed only from the module
web interface. See separate document IN-E-TCPIP-Conf_11 (available in PDF format).

9.2 Ethernet local area network

ROTRONIC devices with an Ethernet (TCP/IP) interface presently use an internal module
manufactured by Digi International (Digi Connect ME for a wired connection, Digi Connect WI-ME
for a wireless connection). Both types of module feature a web interface that is used to configure
the module.

IMPORTANT: The TCP/IP settings of any docking station with Ethernet interface should be
configured to be compatible with the local area network. Detailed instructions are provided
separately in document IN-E-TCPIP-Conf_11 (available in PDF format).