DEVICE SAFEGUARDS

AirChip 3000 Functions

Description

► Device write protection

Used to protect the XB transmitter with a password to prevent
unauthorized digital access by a digital user. Can be configured,
enabled or disabled


PROCESS PROTECTION / PROTECTION OF OTHER DEVICES

AirChip 3000 Functions

Description

► Limit humidity output to 100 %RH

Used to prevent the humidity signal from exceeding
100 %RH when condensation forms on the sensor. Can be
enabled or disabled

► Out-of-limit value alarm

Used to specify the normal range for humidity, temperature and
the calculated parameter depending on the user application.
Can be configured, enabled or disabled

Out-of-limit values trigger a digital alarm which can be also be
seen on the optional display

► Bad sensor alarm

Built-in function. Cannot be disabled

A bad humidity or temperature sensor triggers a digital alarm
which can be also be seen on the optional display

► Fail safe mode

Used to specify a "safe" fixed value for humidity and for
temperature in the event of a sensor failure. Can be configured,
enabled or disabled

4.2 Factory default settings


Notes:

o

Configuration of the XB transmitter by the user and access to its functions requires a PC with the

ROTRONIC HW4 software (version 2.1.1 or higher) installed. Service cable AC3006 or AC3009 is used to
connect the XB service connector to a USB port of the PC.

Configurable Settings

Applicability

Factory default

Unit system (Metric or English)

XB2, XB3

As per ordering code

Analog signal type (4…20 mA or other)

XB3

As per ordering code

Psychrometric calculation

XB2, XB3

As per ordering code

Output 1 parameter, scale and unit

XB2, XB3

As per ordering code (%RH or DP)

Output 2 parameter, scale and unit

XB2, XB3

Temperature, unit as per ordering code


Functions

Applicability

Factory default

Humidity / temperature adjustment

XB2, XB3

Device write protection

XB2, XB3

Disabled

Limit humidity output to 100 %RH

XB2, XB3

Enabled

Out-of-limit value digital / display alarm

XB2, XB3

Disabled