Environmental limits

XB2

XB3

Storage and transit

-50…+70 °C
0…100 %RH, non condensing

Operating limits at electronics

-40 … +85 °C, 0…100 %RH, non condensing

Temperature limits at probe

-100…+200 °C

Maximum humidity at sensor

100 %RH up to 80 °C (176 °F)
75 %RH at 100 °C (212 °F)
45 %RH at 125 °C (260 °F)
15 %RH at 150 °C (302 °F)

Maximum air velocity at probe

40 m/s (7,870 ft /min) – depends on dust filter type

Critical environments

Humidity sensor: as per DV04-14.0803.02 - Critical chemicals

10.2 Dew point accuracy

The XB transmitter can be configured to calculate either the dew point or frost point based on the
measurement of relative humidity and temperature. The accuracy of this conversion varies, depending on the
humidity and temperature conditions as shown in the graph below:

Example: at a temperature of 20 ˚C, a dew point value of -37 ˚C is measured with an accuracy of
± 1.0 ˚C or better.