Procedure for adjusting the XB transmitter with the ROTRONIC HW4 software:

 Use

cable AC3006 to connect the service connector of the XB transmitter to a USB port of a PC with

the HW4 software installed. Note that the ROTRONIC USB driver must be installed on the PC as
explained in the HW4 manual E-M-HW4v2.2-Main.

 Start HW4 on the PC and search for the XB transmitter (HW4 Main Menu Bar > Devices and Groups

> Search for USB Masters).

 After finding the XB transmitter with HW4, expand the device tree to see the XB transmitter functions.

Select Probe Adjustment.

 For further instructions see HW4 manual E-M-HW4v2.2-A2-001

8.4 Cleaning or replacing the dust filter


Depending on the conditions of measurement, the filter should be checked from time to time. Corroded,
discolored or clogged filters should be replaced.

The probe of the XB transmitter has a removable filter cartridge. Simply replace the cartridge (leave the metal
base on the probe).


8.5 Validation of the output signals transmission

If so desired, transmission of the XB transmitter output signals can be validated by using the simulator
function. The HW4 software is required to enable and configure this function. When this function is enabled
the XB transmitter generates fixed digital and analog signals as specified by the user. For instructions see
document E-M-HW4v2.2-F2-009

9

Firmware updates

Firmware updates will be available on the ROTRONIC website for downloading. Firmware files are given a
name that shows both to which device the file applies and the version number of the firmware. All firmware
files have the extension HEX. Procedure for updating the firmware:

 Use cable AC3006 to connect the service connector of the XB transmitter to a USB port of a PC with

the ROTRONIC HW4 software installed. Note that the ROTRONIC USB driver must be installed on
the PC as explained in the HW4 manual E-M-HW4v2.2-Main.

 Copy the firmware update file from the ROTRONIC website to the PC.

 Start HW4 software on the PC and search for the XB transmitter (HW4 Main Menu Bar > Devices

and Groups > Search for USB Masters).

 After finding the XB transmitter, expand the device tree to see the XB transmitter functions. Select

Device Manager. In the Device Manager menu bar select Tools > Firmware Update. For instructions
see document E-M-HW4v2.2-F2-009