E-M-XB-V1_10

Rotronic AG
Bassersdorf, Switzerland

Document code

Unit

Instruction Manual

Document Type

XB Humidity Temperature Transmitter
User Guide

Document title

Page 17 of 27

© 2008; Rotronic AG E-M-XB-V1_10

Terminal block diagram











Measuring humidity or temperature only

Operation of the XB3 does not require both current loops to be closed. When using the XB3 to measure either
humidity only or temperature only, close only the loop that is being used.

Using the ROTRONIC HW4 software, any unused output of the XB3 can be disabled.

Terminals

Description

K1-1: OUT1

Relative humidity or dew point (+)

K1-2: OUT2

Temperature output (+)

K1-3: GND

Ground (tied with other GND)

K2-1: PWR

Power supply: 15…40 VDC (+)
or 12…28 VAC (Phase)

K2-2: GND

Power supply (-) or neutral
(tied with other GND)