E-m-hf5-v2_12 – ROTRONIC HF5 User Manual

Page 19

Advertising
background image

E-M-HF5-V2_12

Rotronic AG
Bassersdorf, Switzerland

Document code

Unit

HygroFlex HF5 Humidity Temperature
Transmitters: User Guide

Instruction Manual

Document Type

Page

19 of 36

Document title

© 2009-2013; Rotronic AG E-M-HF5-V2_12

6.4.3

HF54: 3-wire transmitter, galvanic isolation of the analog outputs


Electrical diagram for voltage outputs

The maximum permissible cable length
can be determined under consideration
of the voltage drop caused by the
current flowing to the devices connected
to the unit. The voltage drop in the cable
depends both on cable resistance and
on the equivalent resistance of the
devices connected in parallel to the unit.
The total resistance connected to each
unit output should be at least 1000
ohms. Cable resistance should not be
more than 1/1000 of the load resistance.


Electrical diagram for current outputs


The maximum permissible cable length,
connecting the unit to other devices, is
determined by the total resistance
resulting from the addition of the cable
resistance and that of the devices
connected in series with the unit. This
resistance should not exceed 500 ohms.





Terminal block diagram (analog outputs only)


U

B

DC

DC

V+

GND

OUT2

OUT1

GND

U

B

DC

DC

V+

GND

OUT2

OUT1

GND