E-M-HF5-V2_12

Rotronic AG
Bassersdorf, Switzerland

Document code

Unit

HygroFlex HF5 Humidity Temperature
Transmitters: User Guide

Instruction Manual

Document Type

Page

29 of 36

Document title

© 2009-2013; Rotronic AG E-M-HF5-V2_12

Main Menu

Menu Items

Selections / Information

Notes

Type

Device type

Name

Device name

User defined

Probe Information

Version

Firmware version

Serial Nbr

Serial number

Address

RS-485 address

Name

Device name

User defined


Other HF5 models:

Main Menu

Menu Items

Selections / Information

Notes

Device Settings

Units

Metric / English

Back Light

Key Press / On / Off

Display backlight mode

Contrast

LC display contrast adjustment

Trend

On / Off

Trend indication on the display

Device Information

Version

Firmware version

Serial Nbr

Serial number

Address

RS-485 address

Type

Device type

Name

Device name

User defined

Probe Information

Version

Firmware version

Serial Nbr

Serial number

Address

RS-485 address

Name

Device name

User defined

SensorTest

Humidity sensor status

Off / Good / SQ-Tuned / Bad

Record

On / Off

Data recording by the probe
(max. 2000 values)

Humidity Adjust

RefValue

Humidity reference value

± 0.1 %RH steps

Acquired

Number of cal. points in probe
memory

<Acquire>

Save cal. point to probe memory

<Delete>

Erases all calibration points

<Adjust>

Effect depends on number of
calibration points

Temperature Adjust

RefValue

Temperature reference value

± 0.1 ˚C steps

<Adjust>

1-point adjustment only (offset)


o

Record: both the recording mode (start / stop and the log interval cannot be changed from the menu and
are as configured with the ROTRONIC HW4 software