Connecting cables – Canon TX-20 Full Auto Tonometer User Manual

Page 38

Advertising
background image

38

4BasicOperation

Connecting Cables

WARNING

• Take the following precautions to avoid risk of fire or electric shock:

- Do not handle the power plug with wet hands.
- Use only the power cord provided with the instrument.

• Connect the cables carefully so there are no problems operating the instrument.

CAUTION

• Always turn off the power and unplug the power cord before connecting the cables of any

external devices to the TX-20.

• Use the computer and other equipment that conform to the system standard IEC 60601-1 or

IEC 60950-1 for the Full Auto Tonometer TX-20. Be sure that the entire system conforms to
IEC 60601-1-1. Be sure to also use an isolation transformer conforming to IEC 60601-1 when a
computer conforming to IEC 60950-1 is used.
Otherwise, electric shock may occur. Ask your sales representative or local Canon dealer to
connect it to a computer.

• If the LAN connector is being used to connect to a network, connect an isolation transformer for

networks between the TX-20 and network devices (HUB etc.) and between network devices.
If this is not done, there is a risk of short circuits or malfunctioning devices. Ask your sales
representative or local Canon dealer to connect it to a network.

1

Make sure that the power switch is off.

2

Connect a USB cable to input patient IDs with USB devices.

Remove the paper from the printer before turning the TX-20 on its side to attach cables. Otherwise
the paper may jam.

• Always turn off the power and unplug the power cord before connecting the USB cable to the

TX-20.

• For details on USB devices for inputting patient ID, contact your sales representative or local

Canon dealer.

USB cable

L-IE-5191C_TX-20.indb 38

2013/07/18 13:10:25