Warning – Canon TX-20 Full Auto Tonometer User Manual

Page 83

Advertising
background image

7Troubleshooting

83

Warning

Code

Cause

Remedy

231

Auto-alignment stopped because the limit of motion (to
right) of the measurement unit was detected during auto-
alignment.

1) Record the message code, contents (including error

codes) and then execute [OK].

2) Be sure that the patient’s forehead and chin are firmly

against the forehead rest and chin rest and then do
measurement again.

232

Auto-alignment stopped because the limit of motion (to
left) of the measurement unit was detected during auto-
alignment.

233

Auto-alignment stopped because the limit of motion (up) of
the measurement unit was detected during auto-alignment.

234

Auto-alignment stopped because the limit of motion
(down) of the measurement unit was detected during auto-
alignment.

235

Auto-alignment stopped because the limit of motion
(forward) of the measurement unit was detected during
auto-alignment.

236

Auto-alignment stopped because the limit of motion
(backward) of the measurement unit was detected during
auto-alignment.

237

Auto-alignment stopped because the set stop position of
the nozzle was detected during auto-alignment.

1) Record the message code, contents (including error

codes) and then execute [OK].

2) Do the following and do the measurement again.

• Press the LIMITER button to reset the position where the

nozzle stops.

• Be sure that the patient’s head and chin are firmly

against the forehead rest.

238

Operation stopped because a button or the operation lever
was operated during auto-alignment.

1) Record the message code, contents (including error

codes) and then execute [OK].

2) Do the measurement again.

241

Measurement stopped because an intraocular pressure
measurement error occurred during full auto-alignment
measurement.

1) Record the message code, contents (including error

codes) and then execute [OK].

2) Do the measurement again.

242

The measured intraocular pressure is over the value set for
the [Upper IOP Warning].

243

The measured intraocular pressure is below the value set
for the [Lower IOP Warning].

244

Auto-alignment stopped because the measurement unit
and the chin rest may be interfering with each other.

1) Record the message code, contents (including error

codes) and then execute [OK].

2) Be sure that the patient’s head and chin are firmly against

the forehead rest.