NeveR use the bed if a zipper slider is bent open or damaged and the zipper is not securely closed.

NeveR use the bed if the zipper pull-tab is missing or broken.

NeveR leave the patient’s bedside until all access panel zippers are securely closed. Quick-

Release Buckles must be connected and closed on all four (

4) access panels before leaving the

patient unattended.

NeveR use the bed if a zipper coil is kinked, misaligned, or has gaps and does not close securely

along the entire length.

Test that all zippers open easily and close securely along the entire length of the zipper.
Inspect zipper coils for any kinks or misalignment. If any are identified, zip and unzip the

zipper. If condition continues,

DO NOt use product. Note the kinks on this sample.

NeveR rip the panels open,

as this will damage the zipper

slider, preventing the zipper

from closing securely.

Check the zippers

Warnings and Precautions

(Continued)

aLWaYs use the zipper pull-tabs to open and close zippers and ensure that zippers are

fully closed.

Opened Up

Damaged pull-tab

Undamaged pull-tab

Crown (Tab) Holder

Body

Pull-Tab

DO NOT use the Posey

Bed if zippers have

open gaps that do not

close securely along

the entire length

Before leaving the patient unattended, apply pressure

with your hands along the entire length of the zipper to

make sure the panel is securely closed and that there

are no gaps or openings along the entire length of

each zipper. Snap the quick-release buckles shut to

ensure all zippers are securely closed, and check by

tugging on each buckle.