Purpose procedure purpose procedure – Posey Bed 8070 User Manual

Page 30

Advertising
background image

30

Posey Bed 8070



Purpose

Procedure

Purpose

Procedure

To keep the Posey Bed clean and disinfected between complete bed and canopy cleanings.

Quick Clean the Posey Bed

machine Washing the Posey Bed

1. Wipe the fabric with cleaning agent

NOte: NeveR use a concentration

of cleaning agent greater than

recommended by the manufacturer.

1. Wash the Canopy on one gentle cycle in cold

water.

NOte: Water should be between 65-85°F

(18-30°C) for approximately 25 minutes.

2. Dry fabric.

3. Clean the Hook and Loop strips.

4. Clean the bed frame.

2. Dry fabric

3. Clean the Hook and Loop strips

4. Follow

transfer Patient into Bed

procedure.

a. Prepare a solution of diluted disinfectant or a

water-based germicide.

b. Use a damp cloth.

c. Rinse with water to remove excess cleaning

agent.

a. Use a mild detergent.

b. Load panels loosely into washing

machine.

c. Rinse thoroughly with water.
Hang dry at room temperature or

machine dry on air-only cycle (no heat).
Use a stiff brush to remove any lint or

foreign materials.

Wipe frame undercarriage with a damp

cloth.

NOte: Ensure cloth is damp, not wet.

Either wipe or air dry before use.

Use a stiff brush to remove any lint or foreign

materials.

aLWaYs wash the canopy between patients to help prevent transmission of infectious disease.

To keep the Posey Bed clean and disinfected between patients.

-

DO NOt use industrial extractor-type laundry equipment, as this may damage the zippers

and buckles.

-

NeveR use bleach, harsh cleaners or solvents.

-

aLWaYs attach hook and loop pieces together. Zip all zippers and connect Quick-Release

Buckles.

-

avOiD motor or bed controls when using water to clean bed frame, as this may damage the

electrical components.

-

uNPLuG the power cord before cleaning bed frame.

stePs

DO this

stePs

DO this