HMI41 AND PROBES
Operating Manual

M210316EN-A

40

10.2

Checking the settings

Turn the HMI41 on and enter the setup mode (see Chapter 8). Make sure that
the following settings are correct:

setting

correct value

probe

AUT

(all probes) or

1 (with HMP41/45) or
2 (with HMP46)

start (*)

1

baud

4.8

seri

E.7.1

calib

def (**)

(*) values 2, 3 and 4 are for calibration cables
(**) when the calib setting is entered, the value is always “no”; factory
settings for calibration correction data can be returned by selecting “def”

If the settings are not correct, change them: in the setup mode, select the
setting with ENTER, change it with buttons ▲ or ▼ and acknowledge with
ENTER. If the error message still appears, write it down and contact Vaisala
or a Vaisala representative for further instructions.