4 ? outputting the hmi41 settings – Vaisala HMP46 User Manual

Page 66

Advertising
background image

HMI41 AND PROBES
Operating Manual

M210316EN-A

62

To change the output, type

CPLAY

, the character you wish to appear between

decimals, the character you wish to use between fields and then <cr>. An
example:

>cplay , <cr>
Desimal separator : ,
Field separator : TAB

example:
1 01:00:00 38,72 21,61 7,01
>

4.2.3

HELP

Outputting available commands and their
contents

Type HELP and press ENTER:

>help
Available commands :
HELP ? PLAY CPLAY
Type HELP <command_name> for more help
>

To see the contents of each command, type

HELP

, command name (e.g.

PLAY

)

and press ENTER:

>help play

Command : PLAY
Purpose : Send recordings from memory to serial port
Usage : PLAY hh:mm <cr>, hh:mm = rec starting time (optional)

if command is used without parameters it uses default setting
>

4.3.4

?

Outputting the HMI41 settings

Type ? and press ENTER:

>?

HMI41 / 2.01
Serial number : A0000000
Output units : metric
Baud P D S : 4800 E 7 1 FDX
Pressure : 1013.25
Auto Off : 5
Probe : 2
Start-up mode : 1
4.th variable : none
>

To exit the terminal session, go to

FILE

menu and select

EXIT

. Confirm that

you wish to quit and then select whether you wish to store the parameters of
this session for future use or not (

SAVE

-

YES

/

NO

).