Connect to the probe in poll mode, Close the connection in poll mode, Reset the probe – Vaisala HMP110 User Manual
Page 52: E 50

User's Guide ______________________________________________________________________
50 __________________________________________________________________ M211060EN-E
Connect to the Probe in POLL Mode
Use the OPEN command to connect to a probe that is in POLL mode.
OPEN [aa]<cr>
where
aa
=
address (0 ... 255)
Example:
open 1
HMP110 1 line opened for operator commands
Close the Connection in POLL Mode
The CLOSE command closes the connection to the probe.
CLOSE<cr>
Example:
close
line closed
Reset the Probe
Use the RESET command to reset the probe. Upon reset or power-up,
the probe enters the serial mode that has been set with the SMODE
command.
RESET<cr>
Example (probe set to serial mode STOP, will output probe model and
software version at reset):
reset
HMP60 / 1.00.0
Example (probe set to serial mode RUN, will start to output measurement
messages at reset):
reset
T= 23.6 'C RH= 20.2 %RH Td= -0.5 'C
T= 23.6 'C RH= 20.2 %RH Td= -0.5 'C
T= 23.3 'C RH= 20.2 %RH Td= -0.8 'C
...