User's Guide ______________________________________________________________________

50 __________________________________________________________________ M211060EN-E

Connect to the Probe in POLL Mode

Use the OPEN command to connect to a probe that is in POLL mode.

OPEN [aa]<cr>

where

aa

=

address (0 ... 255)

Example:

open 1
HMP110 1 line opened for operator commands

Close the Connection in POLL Mode

The CLOSE command closes the connection to the probe.

CLOSE<cr>

Example:

close
line closed

Reset the Probe

Use the RESET command to reset the probe. Upon reset or power-up,
the probe enters the serial mode that has been set with the SMODE
command.

RESET<cr>

Example (probe set to serial mode STOP, will output probe model and
software version at reset):

reset
HMP60 / 1.00.0

Example (probe set to serial mode RUN, will start to output measurement
messages at reset):

reset
T= 23.6 'C RH= 20.2 %RH Td= -0.5 'C
T= 23.6 'C RH= 20.2 %RH Td= -0.5 'C
T= 23.3 'C RH= 20.2 %RH Td= -0.8 'C
...