Repair maintenance, Changing the intercap® sensor (hmp60 and hmp63) – Vaisala HMP110 User Manual

Page 61

Advertising
background image

Chapter 5 ______________________________________________________________ Maintenance

VAISALA _______________________________________________________________________ 59

Repair Maintenance

Changing the INTERCAP® Sensor
(HMP60 and HMP63)

This procedure restores the humidity measurement accuracy of the probe.
No adjustment after the sensor change is needed.

To perform this procedure, you need a new INTERCAP® sensor. It is
also recommended that you replace the filter with a new one. For order
codes, see section Options and Accessories on page 68.

1.

Remove the filter from the probe by turning it counter-clockwise.

2.

There are two sensors under the filter, the INTERCAP® sensor and
a temperature sensor. Identify the INTERCAP® sensor – do not
touch the temperature sensor.

1210-035

Figure 24

INTERCAP® Sensor

3.

Pull out the old INTERCAP® sensor and insert a new one. Refer to
Figure 25 below. Handle the new sensor by the plastic frame.
DO NOT TOUCH THE SENSOR PLATE.

1210-034

Figure 25

Removing the Sensor

4.

Attach a new filter on the probe.