Technical data, Specifications, Performance (hmp60 and hmp63) – Vaisala HMP110 User Manual

Page 65: Relative humidity, Temperature, Dewpoint, Chapter 7, technical data, pr, Technical data on p, Chapter 7

Advertising
background image

Chapter 7 ____________________________________________________________ Technical Data

VAISALA _______________________________________________________________________ 63

CHAPTER 7

TECHNICAL DATA

This chapter provides the technical data of the HMP60 and HMP110
series probes.

Specifications

Performance (HMP60 and HMP63)

Relative Humidity

Measurement range

0 ... 100 %RH

Typical accuracy
temperature range +0 ... +40 °C
0 ... 90 %RH

±3 %RH

90 ... 100 %RH

±5 %RH

temperature range -40 ... 0 °C
and +40 ... +60 °C
0 ... 90 %RH

±5 %RH

90 ... 100 %RH

±7 %RH

Humidity sensor

Vaisala INTERCAP®

Temperature

Measurement range

-40 ... +60 °C

Accuracy over temperature range
+10 ... +30 °C

±0.5 °C

-40 ... +60 °C

±0.6 °C

Dewpoint

Measurement range

-40 ... +60 °C

Typical accuracy
temperature range 0 ... +40 °C
when dewpoint depression* < 15 °C

±2 °C

temperature range -40 ... 0 °C
and +40 ... +60 °C when
dewpoint depression* < 10 °C

±3 °C


*dewpoint depression = ambient temperature - dewpoint

Advertising
This manual is related to the following products: