User's Guide _______________________________________________________________________

162 __________________________________________________________________ M210796EN-G

Table 37

Adjustment and Calibration Commands

Function

Command

Linear corrections on/off

LCI ON/OFF

Entering linear corrections

LCI

Multipoint corrections on/off

MPCI ON/OFF

Entering multipoint corrections

MPCI

Calibration date

CDATE

Opening and Closing the Adjustment Mode

1.

Open the transmitter cover. The buttons needed in adjustment are
on the left-hand side of the motherboard.

2.

If the chemical purge option is available, it should be carried out
always before RH calibration. To start chemical purge press
simultaneously two PURGE push buttons (on the motherboard) for
a few seconds. Red indicator LED flashes with short pulses until
purge is complete (up to 6 minutes).

3.

Press the ADJ button to open the adjustment mode.

4.

Press the ADJ button again to close the adjustment mode.

0508-013

Figure 74

Adjustment and Purge Buttons

The following numbers refer to Figure 74 above:

1

=

Indicator LED

2

=

Adjustment button

3

=

Press the purge buttons simultaneously to start chemical purge
(if available)

Adjustment menu is displayed only when ADJ button (on the
motherboard inside the transmitter) is pressed.