Temperature probe installation, Figure 89, Swagelok installation kit for t-probe – Vaisala PTU300 User Manual
Page 200
Advertising
User's Guide _______________________________________________________________________
198 __________________________________________________________________ M210796EN-G
Temperature Probe Installation
Swagelok installation kit for T-probe includes Swagelok connector with
either ISO1/8" or NPT1/8" thread. Vaisala order codes: SWG6ISO18 or
SWG6NPT18.
0508-016
Figure 89
Swagelok Installation Kit for T-Probe
The following numbers refer to Figure 89 above
1
=
T-probe
2
=
Duct connector
3
=
Swagelok connector
4
=
Ferrules
135 mm
33 mm
18 mm min
max 30 mm
1 2 3
4
P = max 10 bar
T = max 180 ºC