Relative humidity adjustment after sensor change, Using display/keypad, Using serial line – Vaisala PTU300 User Manual
Page 176: Fcrh
Advertising

User's Guide _______________________________________________________________________
174 __________________________________________________________________ M210796EN-G
Relative Humidity Adjustment After Sensor
Change
Using Display/Keypad
When using the optional display/keypad, follow the instructions in
section Using Display/Keypad on page 168 on page but select Adj. for
new RH sensor (instead of 1-point/ 2-point adjustment).
Using Serial Line
After sensor change, carry out the procedure as described in previous
sections. Just replace the CRH command with the FCRH command.
FCRH
Example:
>fcrh
RH
:
1.82 1. ref
?
0
Press any key when ready...
RH
:
74.22
2. ref
? 75
OK
>
The
OK
indicates that the calibration has succeeded.