Example multipoint adjustment procedure, Table 39, Example list for multipoint correction – Vaisala PTU300 User Manual
Page 174
 
User's Guide _______________________________________________________________________
172 __________________________________________________________________ M210796EN-G
Example Multipoint Adjustment Procedure
1.
Connect the PTU300 to a PC. See section Serial Line 
Communication on page 81. Open a terminal program. 
2.
Issue the PURGE command to carry out the chemical purge, and 
wait for six minutes for it to complete. Transmitters without the 
chemical purge feature will not respond to the command. 
 
>purge 
Purge started, press any key to abort... 
 
3.
Press the ADJ button on the component board.
4.
View the currently applied standard user adjustment offset and gain 
using the L command. If offset is zero and gain is one (example 
shown below), no user correction is applied: 
 
>l 
RH offset : 0.00000000E+00 
RH gain : 1.00000000E+00 
T offset : 0.00000000E+00 
T gain : 1.00000000E+00 
 
5.
If some RH offset or gain is applied, write down the adjustment 
and then clear it by writing zero to RH offset and one to RH gain 
using the LI command: 
 
>li 
RH offset : 0.001300000E+00 ? 0 
RH gain : 1.040000000E+00 ? 1 
T offset : 0.00000000E+00 ? 
T gain : 1.00000000E+00 ? 
 
6.
You must now prepare a list of 3 ... 6 correction points using a 
humidity generator or salt solutions (for example, using the 
HMK15 Humidity Calibrator). You must measure the transmitter’s 
RH reading at each point, and make a note of the reference RH 
reading. Here is an example list for a four-point correction: 
Table 39
Example List for Multipoint Correction
Point Number
Transmitter’s Reading
Reference
1
12.2 %RH
11 %RH
2
34 %RH
33 %RH
3
75.5 %RH
75 %RH
4
97.3 %RH
97 %RH