HAPTER

4_____________________________________________________________

COMMISSIONING

V

AISALA

_________________________________________________________________________ 17

The RS485/422 interface of the PTB220 series digital barometers is a
non-isolated two-wire half-duplex interface. See page 93 for quick
reference information on how to use the RS485/422 interface of the
PTB220 series digital barometers.

ECHO Setting the serial bus echo on/off

ECHO x <cr>

where:

x

=

ON or OFF

The ECHO command is used to set or inspect the echoing condition of
the barometer. In OFF mode the barometer does not output the '>'
prompt character.

Examples:

>echo <cr>
Echo

:

ON

>echo off <cr>
Echo

:

OFF

...
echo on

<cr>

(text invisible)

Echo

:

ON

>

FORM Defining the output format

The FORM command is used to set or inspect the output format of the
barometer. See the examples in the next page.

FORM <cr>

The user can define the following fields into the output format:

amount of decimals

give number of decimals before a
quantity. Giving 4.2 before the
pressure quantity outputs a reading
with the following form: 1013.12