SEND Outputting a single message only

SEND <cr>

This command is used to output one message at a time in the STOP
and RUN modes.

Example:

>reset <cr>

PTB220 / 2.02
>send <cr>
1013.25 hPa
>

The user may also use his own previously specified one-message
command (see Chapter SCOM command on page 12).

V Verification message

V <cr>

The V command is used to output the verification message of the
barometer. The verification message outputs the nominal pressure
range of the barometer.

>v <cr>
VAISALA DIGITAL BAROMETER
PTB 220
500.00 to 1100.00 hPa
>

DNOS Self-diagnostics

DNOS <cr>

The DNOS command is used to perform a self-diagnostic routine in
the barometer. If no errors are found, the barometer outputs a PASS
message. In case of error, the barometer outputs an ERR message.
Example:

>dnos <cr>
PASS