Example of closing a barometer:

>close <cr>

line closed

SEND mode

In the SEND mode the barometer will automatically output one
message at power-up or when triggering the barometer on using pin 4
of the barometer. The trigger signal must be a TTL level signal. TTL
HIGH (5 VDC) turns the barometer ON and TTL LOW (0 VDC) turns
the barometer OFF.

Examples of outputting with the prompt (ECHO ON) and without the
prompt (ECHO OFF):

>echo <cr>
Echo

:

ON

>smode send <cr>
Serial mode

: SEND

>reset <cr>
1013.25 hPa
>

>echo off <cr>
Echo

:

OFF

reset <cr>

(text invisible)

1013.25 hPa

Example of outputting a minimum amount of (six) characters by
excluding the <cr> and <lf> characters, the pressure unit and the
decimal point with the FORM command:

>form <cr>
4.2 P " " UUUU #r #n
? 6.0 P <cr>
>unit Pa
P unit

: Pa

>smode send <cr>
Serial mode

: SEND

>echo off <cr>
Echo

:

OFF

reset <cr>

(text invisible)

101325
or

(power-up)

101325