Vaisala PTB220 User Manual

Page 64

Advertising
background image

U

SER

'

S

G

UIDE

_______________________________________________________________________

58 ___________________________________________________________________ M210194

EN

-A

In case of one transducer, the listing looks as follows:

>corr <cr>
Linear adjustments

OFF

Multipoint adjustments

ON

Reading Correction
499.720

-0.070

599.200

-0.080

698.710

-0.010

801.120

-0.010

900.610

-0.030

947.420

-0.020

1000.100

-0.040

1099.580

-0.040

Calibration date

1997-01-01

>

In case of two transducers, the listing looks as follows:

>corr <cr>
Linear adjustments

OFF

Multipoint adjustments

ON

Reading Correction

Reading Correction

499.720

-0.070

499.710

-0.060

599.200

-0.080

599.170

-0.050

698.710

-0.010

698.670

-0.070

801.120

-0.010

801.090

-0.080

900.610

-0.030

900.570

-0.090

947.420

-0.020

947.380

-0.080

1000.100

-0.040

1000.050

-0.090

1099.580

-0.040

1099.500

-0.060

Calibration date

1997-01-01

>

In case of three transducers, the listing looks as follows:

>corr <cr>
Linear adjustments

OFF

Multipoint adjustments

ON

Reading Correction

Reading Correction

Reading Correction

499.720

-0.070

499.710

-0.060

499.640

-0.090

599.200

-0.080

599.170

-0.050

599.160

-0.040

698.710

-0.010

698.670

-0.070

698.640

-0.040

801.120

-0.010

801.090

-0.080

801.070

-0.060

900.610

-0.030

900.570

-0.090

900.560

-0.080

947.420

-0.020

947.380

-0.080

947.390

-0.090

1000.100

-0.040

1000.050

-0.090

1000.060

-0.100

1099.580

-0.040

1099.500

-0.060

1099.540

-0.100

Calibration date

1997-01-01

>