Settings, Baud setting the baud rate, Setting the serial communication parameters – Vaisala PTB210 (serial) User Manual
Page 16: Settings .baud setting the baud rate

User's Guide ______________________________________________________________________
14 __________________________________________________________________ M210669EN-D
Settings
.BAUD Setting the Baud Rate
.BAUD.xxxx<cr>
xxxx =
baud rate (1200 ... 19200)
Example:
.BAUD.9600<cr>
(commands invisible)
.RESET<cr>
NOTE
To activate the new settings, please remember to reset the barometer
with the command .
RESET
<cr>
Setting the Serial Communication
Parameters
The following parameter combinations are available with the PTB210
digital barometers. The factory setting for the parameters is E71.
.E71<cr>
or
.O71<cr>
or
.N81<cr>
The command sets the serial communication parameters of either a
single barometer or all the barometers in.
NOTE
To activate the new settings, please remember to reset the barometer
with the command .
RESET
<cr>